Transcutaneous measurement of hyperbilirubinaemia: comparison of the Minolta jaundice meter and the Ingram icterometer

dc.contributor.authorBEKİROĞLU, GÜLNAZ NURAL
dc.contributor.authorsBilgen, H; Ince, Z; Ozek, E; Bekiroglu, N; Ors, R
dc.date.accessioned2022-03-12T16:55:39Z
dc.date.available2022-03-12T16:55:39Z
dc.date.issued1998
dc.description.abstractThe effectiveness of two different non-invasive transcutaneous bilirubin measurement devices was compared with serum bilirubin levels in 96 healthy newborns. Transcutaneous measurements were obtained with the Minolta Air Shields jaundice meter and the Ingram icterometer and serum bilirubin levels were determined by a direct spectrophotometric method (Bilitron 444). A linear correlation existed between serum bilirubin values and the readings on both the Minolta jaundice meter (r= 0.83) and the Ingram icterometer (r = 0.78). The Kappa coefficient was 0.66. The sensitivity, specificity and positive and negative predictive values were 100%, 56%, 33% and 100% for the Minolta jaundice meter and 100%, 48%, 29% and 100% for the Ingram icterometer, respectively. The high sensitivity and negative predictive value of both devices render them suitable for screening neonatal hyperbilirubinaemia. However, because of its low cost, the Ingram icterometer is preferable to the more complex and expensive Minolta jaundice meter, especially in countries with a high birth rate, such as Turkey.
dc.identifier.doi10.1080/02724936.1998.11747968
dc.identifier.issn0272-4936
dc.identifier.pubmed9924590
dc.identifier.urihttps://hdl.handle.net/11424/226475
dc.identifier.wosWOS:000077338600012
dc.language.isoeng
dc.publisherCARFAX PUBL CO
dc.relation.ispartofANNALS OF TROPICAL PAEDIATRICS
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.subjectTRANS-CUTANEOUS BILIRUBINOMETRY
dc.subjectTERM INFANTS
dc.titleTranscutaneous measurement of hyperbilirubinaemia: comparison of the Minolta jaundice meter and the Ingram icterometer
dc.typearticle
dspace.entity.typePublication
local.avesis.id8c376bc5-8bf4-4f70-b966-20f645e0c611
local.import.packageSS17
local.indexed.atWOS
local.indexed.atSCOPUS
local.indexed.atPUBMED
local.journal.numberofpages4
oaire.citation.endPage328
oaire.citation.issue4
oaire.citation.startPage325
oaire.citation.titleANNALS OF TROPICAL PAEDIATRICS
oaire.citation.volume18