Publication:
Noninvasive technique to monitor the pressure under a cast: A mobile application-friendly bluetooth pressure sensor

dc.contributor.authorGÜNDÜZ, OĞUZHAN
dc.contributor.authorsMahirogullari M., Surucu S., Cerci M. H. , Aydin M., Kayikli A., GÜNDÜZ O.
dc.date.accessioned2022-12-27T05:58:23Z
dc.date.accessioned2026-01-11T19:30:40Z
dc.date.available2022-12-27T05:58:23Z
dc.date.issued2022-01-01
dc.description.abstract© 2022 Mahir Mahirogullari et al.Aim. The purpose of this study was to design a sensor that could monitor the skin-cast contact surface pressure (SCCSP) of a limb under a cast and inform the user via a mobile application when the pressure increases. Methods. In this experimental study, an infant sphygmomanometer cuff was initially placed on the forearm of 10 volunteers. A pressure sensor with a Bluetooth chip was then placed on the volar aspect of the forearm. Short arm plaster was applied with synthetic cast material. The SCCSP under the plaster was measured by the sensor and the measured values were transmitted to a mobile application via a Bluetooth chip. The mobile application processed the data from the chip and converted it to mmHg. Results. Intracompartmental pressure (ICP) values were categorized as 0, 10, 20, 30, 40, 50, 60, and 75 mmHg. The highest SCCSP was 75 mmHg CP, while the lowest was 0 mmHg CP. The correlation coefficient of the mean pressure values was 0.993 (p≤0.001) (SD 0.002, range 0.989-0.997), and there was a significant relationship between ICP and SCCSP values (p≤0.05). Conclusion. We can monitor SCCSP, detect limb swelling, and notify the user via a mobile application by using Bluetooth pressure sensors.
dc.identifier.citationMahirogullari M., Surucu S., Cerci M. H. , Aydin M., Kayikli A., GÜNDÜZ O., "Noninvasive Technique to Monitor the Pressure under a Cast: A Mobile Application-Friendly Bluetooth Pressure Sensor", International Journal of Clinical Practice, cilt.2022, 2022
dc.identifier.doi10.1155/2022/9093612
dc.identifier.issn1368-5031
dc.identifier.urihttps://avesis.marmara.edu.tr/api/publication/54496c78-cb9e-4a56-9d06-87dc5ffb4634/file
dc.identifier.urihttps://hdl.handle.net/11424/284136
dc.identifier.volume2022
dc.language.isoeng
dc.relation.ispartofInternational Journal of Clinical Practice
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectTıp
dc.subjectSağlık Bilimleri
dc.subjectTemel Tıp Bilimleri
dc.subjectMedicine
dc.subjectHealth Sciences
dc.subjectFundamental Medical Sciences
dc.subjectKlinik Tıp (MED)
dc.subjectKlinik Tıp
dc.subjectTIP, GENEL & DAHİLİ
dc.subjectClinical Medicine (MED)
dc.subjectCLINICAL MEDICINE
dc.subjectMEDICINE, GENERAL & INTERNAL
dc.subjectGenel Tıp
dc.subjectGeneral Medicine
dc.titleNoninvasive technique to monitor the pressure under a cast: A mobile application-friendly bluetooth pressure sensor
dc.typearticle
dspace.entity.typePublication

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
file.pdf
Size:
1.62 MB
Format:
Adobe Portable Document Format