Publication:
Clinical gap changes after porcelain firing cycles of zirconia fixed dentures

dc.contributor.authorTÜRKER, ŞEBNEM BEGÜM
dc.contributor.authorsBugurman, Bugurman Burcu; Turker, Sebnem Begum
dc.date.accessioned2022-03-14T10:57:19Z
dc.date.accessioned2026-01-11T08:11:37Z
dc.date.available2022-03-14T10:57:19Z
dc.date.issued2014
dc.description.abstractPURPOSE. The aim of this study was to measure the changes on the marginal and internal adaptation of zirconia based anterior fixed partial dentures after the porcelain firing process. MATERIALS AND METHODS. A total of 34 anterior fixed partial dentures using LAVA CAD/CAM system (3M ESPE, Germany) were applied. Two silicone replicas were obtained: one is obtained before porcelain firing process (initial) and the other is obtained after porcelain firing process (final), followed by the examination under a binocular stereomicroscope. Kruskal Wallis and Wilcoxon Signed Ranks tests were used for the statistical analysis (P<.05). RESULTS. No statistically significant difference was found between initial and final marginal gap values (P>.05). At the internal gap measurements, final marginal area values (59.54 mu m) were significantly lower than the initial marginal area values (68.68 mu m)(P<.05). The highest and the lowest internal gap values were observed at the incisal/occlusal area and at the marginal area, respectively. In. addition, lower internal gap values were obtained for canines than for central incisors, lateral incisors and premolars at the incisal area (P<.05). CONCLUSION. The firing cycles did not affect the marginal gap of Lava CAD/CAM system, but it is controversial for the internal gap.
dc.identifier.doi10.4047/jap.2014.6.3.177
dc.identifier.eissn2005-7814
dc.identifier.issn2005-7806
dc.identifier.pubmed25006381
dc.identifier.urihttps://hdl.handle.net/11424/245581
dc.identifier.wosWOS:000338750900005
dc.language.isoeng
dc.publisherKOREAN ACAD PROSTHODONTICS
dc.relation.ispartofJOURNAL OF ADVANCED PROSTHODONTICS
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectZirconia
dc.subjectResin cement
dc.subjectSurface treatment
dc.subjectMarjinal gap
dc.subjectInternal gap
dc.subjectMARGINAL FIT
dc.subjectDENTAL PROSTHESES
dc.subjectINTERNAL GAPS
dc.subjectIN-CERAM
dc.subjectDISTORTION
dc.subjectCROWNS
dc.titleClinical gap changes after porcelain firing cycles of zirconia fixed dentures
dc.typearticle
dspace.entity.typePublication
oaire.citation.endPage184
oaire.citation.issue3
oaire.citation.startPage177
oaire.citation.titleJOURNAL OF ADVANCED PROSTHODONTICS
oaire.citation.volume6