Publication:
A method for obtaining the pure annihilation peak of Na-22 radioisotope

dc.contributor.authorTAV, CUMALİ
dc.contributor.authorsCeliktas, C; Tav, C
dc.date.accessioned2022-03-12T17:20:08Z
dc.date.accessioned2026-01-11T10:43:30Z
dc.date.available2022-03-12T17:20:08Z
dc.date.issued2005
dc.description.abstractBy means of pulse separation technique, the annihilation gamma ray peak, which has great importance for positron annihilation lifetime spectroscopy (PALS), was discriminated from the superimposed photon spectrum generated by Compton effect, backscattering, and background in the 0.5 MeV energy region. Energy resolution of 4.5% was obtained for the discriminated annihilation peak of Na-22 positron emitter radioisotope. (c) 2004 Elsevier Ltd. All rights reserved.
dc.identifier.doi10.1016/j.radmeas.2004.10.013
dc.identifier.issn1350-4487
dc.identifier.urihttps://hdl.handle.net/11424/228199
dc.identifier.wosWOS:000227823800017
dc.language.isoeng
dc.publisherPERGAMON-ELSEVIER SCIENCE LTD
dc.relation.ispartofRADIATION MEASUREMENTS
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.subjectgamma ray spectroscopy
dc.subjectNa-22 and peak selection
dc.subjectpositron annihilation radiation
dc.subjectPULSE-SHAPE DISCRIMINATOR
dc.titleA method for obtaining the pure annihilation peak of Na-22 radioisotope
dc.typearticle
dspace.entity.typePublication
oaire.citation.endPage572
oaire.citation.issue5
oaire.citation.startPage569
oaire.citation.titleRADIATION MEASUREMENTS
oaire.citation.volume39

Files