Publication:
Dust effect on optically stimulated luminescence dosimetry

dc.contributor.authorsGürdal S.O., Keskin S.S., Tombakoğlu M.
dc.date.accessioned2022-03-15T08:23:41Z
dc.date.accessioned2026-01-11T17:13:10Z
dc.date.available2022-03-15T08:23:41Z
dc.date.issued2017
dc.description.abstractOptically stimulated luminescence dosimetry (OSLD) has been used for dose measurements in many different radiation fields for personal monitoring and medical and industrial applications. One of the most important advantages of the OSLD compared to thermoluminescence dosimetry (TLD) is the light source used to stimulate the crystal. On the other hand, there is a discrimination problem between the light used in a stimulation and the luminescence light obtained as a result of stimulation. To measure the correct dose value, the stimulation and luminescence light have to be discriminated precisely by using different optic filters such as UV blocking, long and short band filters. In addition, the OSL readers are calibrated under fixed conditions (normal operating condition of optic filters, light source, photomultiplier tube, etc). The measured dose values are very sensitive to changes in normal operating conditions. In this work, the dust buildup factor on the optic filters is studied to analyse the response of BeO OSL dosimeter system. The elemental composition of suspended dust was determined by using the literature given for samples obtained from different indoor locations in Turkey and abroad. The light transport algorithm is used to simulate BeO OSL dosimeter system’s response with and without dust buildup by means of Monte Carlo photon transport technique. The Coherent and incoherent scattering of the light, as well as other photon interaction mechanisms, were explicitly modelled in Monte Carlo simulations. The dust buildup effects on OSL spectrum were investigated in detail as a function of dust thickness on the optic filters and elemental composition of the dust. © RAD Conference Proceedings. All rights reserved.
dc.identifier.doi10.21175/RadProc.2017.12
dc.identifier.issn24664626
dc.identifier.urihttps://hdl.handle.net/11424/248449
dc.language.isoeng
dc.publisherRAD Association
dc.relation.ispartofRAD Conference Proceedings
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectDosimetry
dc.subjectDust effect
dc.subjectMonte Carlo simulation
dc.subjectOptically Stimulated Luminescence Dosimetry (OSLD)
dc.titleDust effect on optically stimulated luminescence dosimetry
dc.typeconferenceObject
dspace.entity.typePublication
oaire.citation.endPage58
oaire.citation.startPage55
oaire.citation.titleRAD Conference Proceedings
oaire.citation.volume2