Publication:
Numerical analysis of a dental implant system in three-dimension

dc.contributor.authorsEkici, B
dc.date.accessioned2022-03-12T16:58:53Z
dc.date.accessioned2026-01-10T17:19:01Z
dc.date.available2022-03-12T16:58:53Z
dc.date.issued2002
dc.description.abstractOne of the important problems in dental implant systems is screw loosening. In order to prevent this loosening a preload is applied to the retaining screw. However, this measure has not been able to eliminate its occurrence. In this study the effect of a washer in a Branemark-type implant on the loosening conditions of the retaining screw was investigated using a finite element simulation in three-dimensional models. The simulation indicated that a washer has an important affect against loosening. This also increases the flexibility of implant system against the deformation. The application of a washer is easy and cheap. (C) 2002 Elsevier Science Ltd. All rights reserved.
dc.identifier.doi10.1016/S0965-9978(01)00056-4
dc.identifier.eissn1873-5339
dc.identifier.issn0965-9978
dc.identifier.urihttps://hdl.handle.net/11424/227119
dc.identifier.wosWOS:000173818100005
dc.language.isoeng
dc.publisherELSEVIER SCI LTD
dc.relation.ispartofADVANCES IN ENGINEERING SOFTWARE
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.subjectthree-dimensional contact modeling
dc.subjectimplant system
dc.titleNumerical analysis of a dental implant system in three-dimension
dc.typearticle
dspace.entity.typePublication
oaire.citation.endPage113
oaire.citation.issue2
oaire.citation.startPage109
oaire.citation.titleADVANCES IN ENGINEERING SOFTWARE
oaire.citation.volume33